What is the operating temperature range for the Faro arm?

The operating temperature range for the platinum arm is 10C-40C. This is a broad range, however for the arm to function properly a stable environment is required. The temperature should not vary by more than 5 Celsius degrees in 5 minutes. If the arm is moved to a new environment, it should be allowed to adjust to that temperature environment for at least two hours before using the arm.

Rating:Rating of 0 Stars1 Votes
Was this answer helpful?YesNo